ACNET console
Instructions to reset the clcsend program :


- rlogin b0clc01
> moduleShow (shows Modules)
> i (shows processes - get task name for clcsend)
> td [task name, i.e. t4]
OR
> clcstop (this ends the clcsend directly)

To start the program again:
> sp clcsend

------------------------------------------------------ Instructions to run the ACNET console :

- telnet cns55 nomerot/(ask Andrei for psw)
> cnsrun (this will pop up a lot of windows)
    - to quit :
            go to the Utilities window to 'Shutdown console'
    - to see luminosity/losses :
           go to any Parameters window type in the left upper corner 'C65'
           this gets you to the 'CDF Luminosity' page where the following parameters are defined :

            B0LUM - total instanteneous luminosity
            B0LUM[1],...,B0LUM[36] - 36 inst.lum. by bunch
            B0PLOS - total proton losses
            B0PLOS[1],...,B0PLOS[36] - 36 proton losses by bunch
            B0ALOS - total antiproton losses
            B0ALOS[1],...,B0ALOS[36] - 36 antiproton losses by bunch

   - you can view subpages 1-7 which are preset to view all of above or
   - you can do time plots of any of them defining X and Y variables and their limits (upper lines in the window)

   - so far rates [Hz] are transmitted in the following order :

       B0LUM[1] - BSC1W.1
       B0LUM[2] - BSC1W.2
       B0LUM[3] - BSC1W.3
       B0LUM[4] - BSC1W.4
       B0LUM[5] - (BSC1W.1).or.(BSC1W.2)
       B0LUM[6] - (BSC1W.3).or.(BSC1W.4)
       B0LUM[7] - ((BSC1W.1).or.(BSC1W.2)).and.((BSC1W.3).or.(BSC1W.4))
       B0LUM[8] - NIM pulser rate ~50 Hz
       B0LUM[9] - CDF_CLK rate ~7.6 MHz